ISO/IEC 17025-2017 Accredited Calibration

Com-Power is proud to announce that we are now ISO-17025 Accredited for Calibration. 

We are currently accredited for:

  • Line Impedance Stabilization Networks (LISNs)
  • Impedance Stabilization network (ISNs)
  • Coupling/Decoupling Networks (CDNs) and 150Ω to 50Ω Adapters
  • Current Probes and BCI Probes
  • Preamplifiers

Please click here for our certificate and scope:

Com-Power Corporation Certificate Number AC-2894

 

ISN
Impedance
Phase
Voltage Division Factor
Isolation (Decoupling Factor)
Longitudinal Conversion Loss

150 kHz to 230 MHz
(0.1 to 210) Ω
(0 to 360) °
(-15 to 0) dB
(-80 to 0) dB
(-85 to -35) dB

1.8 % of reading
1.2 °
0.27 dB
2.4 dB
0.7 dB

CISPR 22, CISPR 32,
CISPR 16-1-2,
Agilent 4395A Network
Spectrum Impedance
Analyzer w/87511A
S-Parameter Test Set,
Agilent 4294A Precision
Impedance Analyzer, Cal Kit

LISN
Impedance
Phase
Insertion Loss
Isolation (Decoupling Factor)

9 kHz to 400 MHz
(0.1 to 210) Ω
(0 to 360) °
(-10 to 0) dB
(-80 to 0) dB

1.8 % of reading
1.2 °
0.27 dB
2.4 dB

CISPR 22, CISPR 32,
CISPR 16-1-2, ANSI C63.4,
Mil-Std-461, RTCA DO-160,
Agilent 4395A Network Spectrum
Impedance Analyzer w/87511A
S-Parameter Test Set,
Agilent 4294A Precision
Impedance Analyzer, Cal Kit

Current Probes & BCI Probes
Insertion Loss (Transfer Impedance)

9 kHz to 400 MHz
(-70 to 0) dB
400 MHz to 1 GHz
(-70 to 0) dB

0.28 dB
0.79 dB
CISPR 16-1-2, Mil-Std461,
RTCA DO-160,
Agilent 4395A Network
Spectrum Impedance
Analyzer w/87511A
S-Parameter Test Set,
Agilent 8722ES S-Parameter
Vector Network Analyzer

CDN & 50Ω-150Ω Adapters
Impedance
Phase
Voltage Division Factor
Isolation (Decoupling Factor)
Adapter Insertion Loss

150 kHz to 230 MHz
(0.1 to 210) Ω
(0 to 360) °
(-20 to 0) dB
(-80 to 0) dB
(-11 to -8) dB
1.8 % of reading
1.2 °
0.27 dB
2.4 dB
0.15 dB
CISPR 16-1-2,
IEC/EN 61000-4-6,
Agilent 4395A Network
Spectrum Impedance
Analyzer w/87511A S-Parameter
Test Set, Agilent 4294A Precision
Impedance Analyzer, Cal Kit
RF Preamplifiers, Amplifiers
Gain
100 Hz to 500 MHz
(0 to 60) dB
50 MHz to 40 GHz
(0 to 60) dB
0.27 dB
0.47 dB
Agilent 4395A Network Spectrum
Impedance Analyzer,
87511A SParameter Test Set,
Agilent 8722ES S-Parameter
Vector Network Analyzer

 

Current Offers